← Spitalul Orasenesc Bolintin Vale
fier ferozina
DA12616964 · published 2017-07-03 · completed 2017-07-04 · from Biosystems Diagnostic S.R.L. (Bucuresti)
93 leiawarded value
—estimated value
33696500-0CPV code
The quantity is not published in the open data. The MS/SEAP export has only the title and value; quantity and unit price appear only in the purchase's SICAP record, which you can look up on e-licitatie.ro.
Description from SEAP
CONTINUT SI COMPOZITIEA.Reactiv. 4 x 40 ml. Clorura de guanidinium 1.0 mol/l, tampon acetat 0.4 mol/l, pH 4.0B.Reactiv. 4 x10 ml. Ferrozina 8 mmol/l, acid ascorbic 200 mmol/lS.Fier standard. Standard lichid.Concentratia este indicata pe eticheta.Concentratia este trasabila la Standard Reference Material 937(National Institute of Standards and Technology, USA)ECHIPAMENT ADITIONALSer sau plasma heparinizata, recoltate prin proceduri standard.CARACTERISTICI METROLOGICE•Limita de detectie:4 µg/dl = 0.71 µmol/l•Limita linearitatii:1000 µg/dl = 179 µmol/l. Pentru valori mai mari diluati proba1/2 cu apa distilata si repetati masurarea•Repetabilitate ( in cadrul unei serii de masurari)Concentratie medie CV n103 µg/dl=18.4 µmol/l 2.2% 20305 µg/dl= 54.6 µmol/l 0.7% 20•Reproductibilitate( de la o serie de masurari la alta).Concentratie medie CV n103 µg/dl=18.4 µmol/l 2.9% 20305 µg/dl= 54.6 µmol/l 2.0% 20•Sensibilitate: 88 mA*dl/µg = 4.86 mA*l/µmol •Veridicitate: rezultatele obtinute cu acest reactiv nu au aratat diferente sistematice prin comparatie cu reactivi de referinta.(observatia 3).Detalii ale experimentelor comparative sunt disponibile la cerere.•Interferente: bilirubina (<20 mg/dl) si lipemia (10 g/l) nu interfera.Hemoliza interfera.Alte medicamente si substante pot cauza interferente5.